94 880 Technical/Service Manual
10.0 Appendix
10.1 Error Messages
The 880 indicator provides a number of error messages. When an error occurs, the message is shown on the
indicator display. Error conditions can also be checked remotely by using the XE EDP command as described in
Section 10.3.
10.1.1 Displayed Error Messages
The 880 provides a number of front panel error messages to assist in
problem diagnosis. Table 10-1 lists these
messages and their meanings.
Table 10-1. 880 Error Messages
Error Message Description Solution
Over range Check for improper load cell wiring, conf
iguration, calibration, scale
hardware problems.
Under range
A/D out of range
Or if using local/remote (serial scale) -
l
oss of serial scale data.
CFGERR Configuration error on power up if
ther
e was an error loading
configuration
Press the Enter key to reboot the indicator.
ERROR Internal program error Check configuration.
HWFERR Hardware failure error
• on failure to write to the EEPROM
• any error (except for a battery error
or an accumulation over range error)
when exiting the menu
Press the Enter key to reboot the indicator.
LOBATT The low battery error flashes every 30
seconds when
the battery is low.
Replace the battery.
NOTARE Tare is prevented because of
r
egulatory mode settings, the
configuration of the TAREFN
parameter, motion on the scale, etc.
Change regulatory mode settings or the TAREFN parameter.
RANGE A numeric value entered in
c
onfiguration is out of the acceptable
range. The error is displayed
momentarily – then parameter being
edited is displayed so the value can
be corrected.
Re-enter a value that is in range f
or the parameter being edited.
NO ZERO Zero is prevented (due to regulatory
mo
de settings, motion on the scale,
zero range settings)
Check zero settings and for motion.
¯ ¯ ¯ ¯ ¯ ¯
_ _ _ _ _ _
(center dashes)