Rice-lake Counterpart Configurable Counting Indicator - User Manual do Utilizador Página 1

Consulte online ou descarregue Manual do Utilizador para Balanças Rice-lake Counterpart Configurable Counting Indicator - User. Rice Lake Counterpart Configurable Counting Indicator - User Manual Rev. D - Version 2.4 User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 112
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes

Resumo do Conteúdo

Página 1 - User Manual

CounterpartCounting Scale IndicatorVersion 2.4User Manual®118677 Rev D

Página 2

4 Counterpart Counting Scale IndicatorFigure 1-1. Numeric KeypadAdditional Keypad SymbolsAdditional symbols can be accessed by pressing the following

Página 3 - Contents

94 Counterpart Counting Scale Indicator9.6 Digital FilteringThe Counterpart uses averaged digital filtering to reduce the effect of vibration on weigh

Página 4

Appendix 959.6.3 Setting the Digital Filter ParametersFine-tuning the digital filter parameters greatly improves indicator performance in heavy-vibrat

Página 5

96 Counterpart Counting Scale Indicator9.7 USB Keyboard InterfaceThe USB interface option board provides a type-A connection for a USB keyboard interf

Página 6

Appendix 97Where:<STX> STX character<P> Polarity<W7> Seven characters of net data with decimal pt<U>

Página 7 - About This Manual

98 Counterpart Counting Scale Indicator9.10 Regulatory Mode and Legal-for-Trade (LFT)NoteWhen REGULAT is set for NONE or INDUST there are no restricti

Página 8 - 1.0 Introduction

Appendix 999.10.2 Legal for Trade and Totalization ModeWhen in Legal for Trade mode (REGULAT = NTEP, OIML, or CANADA) and the unit is in Totalization

Página 9 - 1.4 Keypad Functions

100 Counterpart Counting Scale Indicator3. Install jumpers on JP5 and JP6. See Table 2-5 on page 16.4. Plug in power to the Counterpart and press the

Página 10 - 1.6 Annunciator Symbols

Appendix 1019.12 SpecificationsTable 9-8. SpecificationsPart # Description Weighing Resolution Platter DimensionsSingle Channel118788 CP Indicator Con

Página 11 - 1.9 Indicator Operations

102 Counterpart Counting Scale Indicator9.13 SpecificationsPOWER:Power source:Input: 100-240VAC, 47-63 Hz, 5 watts, US power cordOutput: 9VDC o

Página 12

Appendix 103Counterpart Limited WarrantyRice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by a Distrib

Página 13 - Important

Introduction 51.7 Softkey SetupSoftkeys offer additional ways to access features associated with the Counterpart. To access the softkey setup paramete

Página 14

104 Counterpart Counting Scale IndicatorFor More InformationLiterature• Counterpart Sales Literature, PN 125545• Counterpart CD with WeighVault, PN 12

Página 16 - 2.6 Load Cell Wiring

230 W. Coleman St. • Rice Lake, WI 54868 • USAU.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0) 88 2349171w

Página 17 - 2.7 Wire Specifics

6 Counterpart Counting Scale IndicatorSet Date and Time1. If the softkey is enabled, push the softkey. The current value is displayed.2. To change, us

Página 18 - 2.8 Cable Grounding

Installation 72.0 InstallationCounterpart is available in different configurations from just the indicator display to the display and counting scale c

Página 19 - 2.10 CPU Board Removal

8 Counterpart Counting Scale Indicator2.2 Scale Base Assembly (if purchasing the scale base separate from indicator)ImportantDo not turn the scale ups

Página 20

Installation 92.2.3 LevelingSelect a location for the Counterpart that is reasonably level and free of vibrations and air currents. Adjust the four co

Página 21 - 2.13 Installing Option Cards

10 Counterpart Counting Scale Indicator5. Push the plastic bushing all the way into the enclosure to make a tight seal in the hole. Push bushing all t

Página 22 - Figure 2-14. CPU Board

Installation 112.6.1 Dual Channel Units (S-XD Base)For units that have dual channels, connect cable from the 14 pin connector to J1 and J2. If not usi

Página 23

12 Counterpart Counting Scale Indicator2.8 Cable GroundingExcept for the power cord, all cables should be grounded against the scale enclosure. Do the

Página 24

Installation 132.8.2 Digital I/OThe Digital I/O can be configured as either digital inputs or digital outputs as determined by the Digital I/O menu (s

Página 26

14 Counterpart Counting Scale Indicator2.11 CPU Board Battery ReplacementCAUTIONRisk of explosion exists if battery is replaced with incorrect type. D

Página 27 - Counterpart Dimensions

Installation 157. Run cable over to power plug J12 on the Counterpart CPU board.Figure 2-13. Connect Power Plug onto J12 on CPU Board8. Unplug existin

Página 28 - 3.0 Configuration

16 Counterpart Counting Scale IndicatorWhen installation is complete, reassemble the enclosure as described in Section 2.9 on page 13LEDLEDJP1 JP2JP3

Página 29 - 3.3 Revolution Configuration

Installation 174. Press down firmly to seat the option card in the CPU board connector.D5 LEDD6 LEDD1 LED Figure 2-15. Installing the USB Option Card5

Página 30 - 3.4 ID Menu

18 Counterpart Counting Scale Indicator1. Stand scale base on its side to attach the bracket to the scale using the screws provided in the bracket kit

Página 31 - 3.5 Audit Menu

Installation 192.15 Indicator Replacement Parts and Assembly DrawingsRef Number Part Number Description (Quantity)1 118434 Overlay, membrane switch pa

Página 32 - 3.6 Calibration Menu

20 Counterpart Counting Scale IndicatorFigure 2-22. Counterpart Assembly and Components

Página 33 - 3.7 Setup Menu

Installation 21Figure 2-23. Counterpart DimensionsFigure 2-24. Counterpart Rear and Side Dimensions

Página 34 - 3.7.1 Scale Menu

22 Counterpart Counting Scale Indicator3.0 ConfigurationThe Counterpart scale can be configured using a series of menus accessed through the scale fro

Página 35

Configuration 233.3 Revolution ConfigurationThe Revolution configuration utility provides the preferred method for configuring the Counterpart indicat

Página 36

Contents iContents1.0 Introduction...

Página 37 - Scale Calibration Menu

24 Counterpart Counting Scale Indicator3.4 ID MenuThe ID menu displays parameters for registers from 1 to 150. Sub-parameters are shown below.VALUE VA

Página 38 - Scale Format Menu

Configuration 253.5 Audit MenuThe audit menu accesses audit trail support. It provides tracking information for configuration and calibration events.

Página 39 - Scale Format Menu Parameters

26 Counterpart Counting Scale Indicator3.6 Calibration MenuSee Section 4.0 on page 55 for calibration procedures. The Calibration menu can be protecte

Página 40 - 3.7.2 Feature Menu

Configuration 273.7 Setup MenuThe setup menu allows: • Configuration of scale, features, serial port, Ethernet, print format, and digital inputs and o

Página 41 - UNITWTUPDATE

28 Counterpart Counting Scale Indicator3.7.1 Scale MenuCalibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-6

Página 42 - Feature Count Menu

Configuration 29Parameter Choices DescriptionScales 1 and 2 Allows configuration and calibration of each scale.Scale 3 Allows configuration and calibr

Página 43

30 Counterpart Counting Scale IndicatorDFTHRH NONE2D5D10D20D50D100D200D250DDigital filter cutout threshold. Specifies the filter threshold, in display

Página 44 - Feature Region Menu

Configuration 31Scale Calibration MenuCalibration can be performed in two places within the menu: the CALIBR menu shown in Figure 3-6 on page 26 is an

Página 45 - SOFTKEYS Menu

32 Counterpart Counting Scale IndicatorScale Format MenuFORMATSECTERPRIMRYGOFFKGLBOZLBKGGOZOFF5D1D2DOFFLBOZKGGUNITS DECPNT DSPDIV888888888888008888880

Página 46 - Regulate Industry Menu

Configuration 33DECPNT 8888800888888088888888.88888888.88888888.88888888.88888888.88888888.8Allows you to place the decimal point position. Use the ID

Página 47

ii Counterpart Counting Scale IndicatorRice Lake continually offers web-based video training on a growing selection of product-related topics at no co

Página 48 - 3.7.3 Serial Menu

34 Counterpart Counting Scale Indicator3.7.2 Feature MenuTARE RND1COUNTREGIONUIDSOFTKEYS STNDBY0RECALL0NOFFDISPLAY KEYBEEPOFFONCONTRASTVALUEIMAGEVALUE

Página 49

Configuration 35 Parameter Choices DescriptionCOUNT NEGCOUNTSAMPLEQTYINSFSMPLLOTUPDTSCLCHGNEWITEMXFRSamplDISPACCUNITWTUPDATEUTWTBASEDSP ModeRemovESPSe

Página 50 - 3.7.4 Ethernet Menu

36 Counterpart Counting Scale IndicatorFeature Count MenuCOUNTNEGCOUNT0.1%0.2%0.0%INSFSMPLSAMPLQTY102550100LOTUPDTOFFONNEWITEMONOFFSCLCHGCHKSTBLEONOFF

Página 51 - 3.7.5 Print Format Menu

Configuration 37LOTUPDT OFFONOFF — Changing or deleting the lot value in operation mode does not update the stored copy of the currently loaded ID rec

Página 52 - 3.7.6 Scale Setpoints Menu

38 Counterpart Counting Scale IndicatorFeature Region MenuTFORMTT SEPDFORMT D SEPTIMENTEPCANADAINDUSTGROSSBRUTTODOTCOMMANONEOIMLCOLONCOMMASLASHDASHSEM

Página 53 - Sub-menu

Configuration 39SOFTKEYS MenuFigure 3-14. Softkey Menu LayoutDate sub-menu DFORMT MMDDY4DDMMY4Y4MMDDY4DDMMMMDDY2DDMMY2Y2MMDDY2DDMMSets the date format

Página 54

40 Counterpart Counting Scale IndicatorRegulate Industry MenuOnly displayed if INDUST is selected from the REGULAT parameterREGULATINDUSTDISPLYSCALESN

Página 55

Table 3-14. Regulate Industry Menu ParametersParameter Choices DescriptionSNPSHT DISPLYSCALEDisplay or scale weight source.ZTARE NO YESRemove tare on

Página 56 - 3.7.7 DIG I/O Menu

42 Counterpart Counting Scale Indicator3.7.3 Serial MenuONOFFCRCR/LF217ODD7EVEN8NONESERIALPORT1PORT3 OPTCARDCMDSCANNERSCALE IND SCALEBAUD RESPONSEBITS

Página 57 - 3.8 Test Menu

Configuration 43OPTCARD NONEFIBERETHERNETWI-PORTRESERVEDUSBRS232-432KEYBOARDOption card parameters. Indicates which connection can be selected when us

Página 58 - 3.9 Display Menu

Contents iii7.0 Print Formatting ...

Página 59 - SETPT 2 - 6

44 Counterpart Counting Scale Indicator3.7.4 Ethernet MenuONOFFETHERNETDHCPIPADDRESSNETMASK DFLGTWY DNSPRI DNSSECLCLHSTNMPORTCOUNTERPARTMAC REMOTEPT R

Página 60

Configuration 453.7.5 Print Format MenuPFORMTPORTPORT 1FMT<x><x>GFMT NFMTCFMTPALFMTSAME AS GFMT<x><x>HDRFMT1<x><x>

Página 61 - 4.0 Calibration

46 Counterpart Counting Scale Indicator3.7.6 Scale Setpoints MenuSETPT 1SETPTSBATCHINGOFFAUTOMANUALSETPT 1-6Same as SETPT 1SP CFGOFFGROSS NET PIECECNT

Página 62 - 4.4 EDP Command Calibration

Configuration 47Gross, Net, Piecent, %REL GROSS NET PIECECNT %RELAPPLIES TO GROSS, NET, PIECECNT, & %RELSOURCELIST OF SCALESVALUEnumberTRIPLOWERHI

Página 63 - Calibration

48 Counterpart Counting Scale IndicatorSetpoint 1-6 submenuOFF Value = numberGROSSNETPIECECNT%RELValueSourceTripBandvalHysterPreactPrevalPreadjPrestab

Página 64 - 4.6 More About Calibration

Configuration 49PREACT OffOnLearnAllows the digital output associated with a setpoint to shut off before the setpoint is satisfied to allow for materi

Página 65 - 5.0 Scale Operations

50 Counterpart Counting Scale Indicator3.7.7 DIG I/O MenuDIGIODIO 1 DIO 2 DIO 3 DIO 4OFFPRINTUNITSTARENT/GRSDSPACCCLEARZEROCLRCNSAMPLESOFTKEY 1-4OUTPU

Página 66 - = 0.098%

Configuration 513.7.8 Version MenuVERSSOFTWR REGV 2.0 LR, V.1.00Figure 3-22. Version Menu LayoutTable 3-18. Version Menu ParametersParameter Choices D

Página 67

52 Counterpart Counting Scale Indicator3.9 Display MenuThere are three display settings on the Counterpart. 0-100%50NegativePositiveID DISPLAYTESTSETU

Página 68 - Figure 5-1. ID Menu

Configuration 533.10 Setpoints – Weigh Mode Parameter MenuSet up of setpoints is allowed while in the weigh mode. While setting up setpoints, the Acce

Página 69 - 5.5.2 Using a Stored ID

iv Counterpart Counting Scale Indicator

Página 70 - 5.5.4 Clearing ID Codes

54 Counterpart Counting Scale IndicatorSetpoint 1-6 sub-menuGROSSNETPIECECNT%RELValueSourceTripBandvalHysterPreactPrevalPreadjPrestabPcountRelnumBatch

Página 71 - 5.6 WeighVault

Calibration 554.0 CalibrationThe Counterpart can be calibrated using the front panel, EDP commands, or Revolution®. NoteCalibration can be performed i

Página 72 - 5.6.2 Using WeighVault

Note56 Counterpart Counting Scale Indicator If you do not want to calibrate span, press Menu to exit.6. After the Calibrate softkey is pressed, the

Página 73 - 5.7 Totalization Counts

Calibration 575. Up to five linearization points can be calibrated between the zero and span calibration values. Use the following commands to set and

Página 74 - 5.8 Parts Reduction Counts

58 Counterpart Counting Scale Indicator4.6 More About CalibrationThe following provides additional information about compensating for environmental fa

Página 75 - 5.10 Batch Operations

Scale Operations 595.0 Scale OperationsThe following contains detailed operator instructions for Counterpart, including instructions on how to enter t

Página 76 - DIGITAL I/O

60 Counterpart Counting Scale Indicator5.3 Toggling Between Gross and NetTo toggle between net and gross weight, a tare value must be entered into the

Página 77 - 5.12 USB Keyboard

Scale Operations 615.4.1 Unit Weight Operation by SamplingUnit weight operation by sampling is accomplished by placing a 10 piece sample on the scale

Página 78 - 6.0 Serial Commands

62 Counterpart Counting Scale IndicatorCounting Out of a Full Container — See Total Amount Remaining in the ContainerTo carry out this operation you m

Página 79 - 6.1.3 Reporting Commands

Scale Operations 635.5.1 Setting an ID Using the Normal Mode1. Press the MENU SETUP key MENUSETUPto access this menu.2. Press to ID and then press

Página 80

1About This ManualThis manual is for trained and qualified installers of counting scales and represents the correct, safe and recommended methods for

Página 81

64 Counterpart Counting Scale Indicator5.5.4 Clearing ID Codes1. Press the Menu Setup key.2. Toggle through menu items to ID.3. Select Reg 1-150 to cl

Página 82

Scale Operations 658. To download the ID to Counterpart, connect Revolution, and select Communications and either Download Configuration or Download S

Página 83

66 Counterpart Counting Scale IndicatorOn the CounterpartUse the following steps to set up WeighVault on the Counterpart side.1. Enter menu settings a

Página 84 - 6.1.7 Unique Commands

Scale Operations 675.7 Totalization CountsCounterpart has a totalization function that allows you to totalize the quantity of several weighings togeth

Página 85 - 7.0 Print Formatting

68 Counterpart Counting Scale Indicator5.8 Parts Reduction CountsParts reduction can also be done by using the minus key while the scale is in the wei

Página 86

Scale Operations 695.10 Batch OperationsSoftkeys can be configured to allow the operator control of batch operations from the Counterpart front panel

Página 87

70 Counterpart Counting Scale Indicator.REDWHITEBLACKREDBLACKSTOP/START MUSHROOM SWITCHABORT/RUN/START SWITCHABORTSTART3412NO NC NCNO NOJ3+5VDCGNDDIO1

Página 88 - Table 7-3. H1, H2 Formats

Scale Operations 715.11 Connecting a Barcode ScannerThe Counterpart will accept a barcode scanner connected to the unit. In order to use a scanner, yo

Página 89 - 7.2 Customizing Print Formats

72 Counterpart Counting Scale Indicator6.0 Serial CommandsThe Counterpart can be controlled by a PC or remote keyboard connected to an indicator seria

Página 90 - 7.2.4 Print Label Examples

Serial Commands 736.1.2 ID CommandsUp to 150 codes can be entered under ID commands.Table 6-2. ID CommandsCommand FunctionID.CODE#n ID code ID.DESC#n

Página 91 - Gross Label Format

2 Counterpart Counting Scale Indicator1.0 IntroductionThe Counterpart offers practical solutions for a full range of precision counting applications.

Página 92 - Total Label Format

74 Counterpart Counting Scale Indicator6.1.4 Clear and Reset CommandsThe following commands can be used to clear and reset the Counterpart:RS: Reset s

Página 93 - Accum Label Format

Serial Commands 75SC.PRI.DSPDIV#n Primary units display divisions 1D, 2D, 5DSC.PRI.UNITS#n Primary units lb, kg, g, oz, OFFSC.ACCUM#n Accumulator enab

Página 94 - 8.1 Enclosure Disassembly

76 Counterpart Counting Scale IndicatorSTR.POS#p Custom stream identifiers None, Space, +STR.NEG#p None, Space, -STR.PRI#p 8 alphanumeric charactersST

Página 95 - - - - - -

Table 6-7. PFORMT Serial CommandsCommand Description ValuesGFMT.FMT Gross print format string See Section 7.0 on page 79 for information about demand

Página 96 - 9.2.1 Using the P EDP Command

78 Counterpart Counting Scale Indicator6.1.6 Normal Mode CommandsThe normal mode print commands transmit data to the serial port on demand in either s

Página 97 - <LF>

Print Formatting 797.0 Print FormattingThe Counterpart provides nine print formats, GFMT, NFMT, TOTALFMT, PALFMT, CFMT, ACCFMT, SPFMT, HDRFMT1, and HD

Página 98

80 Counterpart Counting Scale IndicatorCommand Description Supported Ticket Formats<G> Gross weight, current scale GFMT, NFMT, TOTALFMT, PALFMT,

Página 99

Print Formatting 81<WL> Suppress units. Toggle weight data format (formatted/unformatted)** GFMT, NFMT, TOTALFMT, PALFMT, CFMT, ACCFMT HDRFMT1,

Página 100 - 9.6 Digital Filtering

Table 7-2. Default Print FormatsFormat Default Format String Used WhenNFMTN<NL>A30,75,0,4,1,1,N,"Gross: <G>"<NL>B30,115,0,

Página 101 - 9.6.4 Audit Trail Support

Print Formatting 837.2 Customizing Print FormatsThe following sections describe procedures for customizing the GFMT, NFMT and CFMT formats using the E

Página 102 - 9.8 Serial Scale Interface

Introduction 31.4 Keypad Functions Table 1-2. Keypad Functions Key Function POWERTurns the Counterpart unit on/off.NoteIf the PC1 jumper is set

Página 103 - 9.9 Regulatory Mode Functions

84 Counterpart Counting Scale Indicator7.2.3 Using Revolution®The Revolution configuration utility provides a print formatting grid with a tool bar. T

Página 104

Print Formatting 85Count FormatFigure 7-3. Count Format Label ExamplePallet Label FormatFigure 7-4. Pallet Format Label ExampleGross Label FormatFigur

Página 105 - 9.11 Updating Firmware

86 Counterpart Counting Scale IndicatorSetpoint FormatFigure 7-6. Setpoint Label FormatTotal Label FormatFigure 7-7. Total Format Label Example

Página 106 - Figure 9-4. Revolution Screen

Print Formatting 87Gross/Tare/Net Label FormatFigure 7-8. Gross/Tare/Net Label FormatAccum Label FormatFigure 7-9. Accum Label Format

Página 107 - 9.12 Specifications

88 Counterpart Counting Scale Indicator8.0 WLAN Installation InstructionsNoteBefore installing this option, contact your IT administrator to obtain ne

Página 108 - 9.13 Specifications

Appendix 899.0 Appendix9.1 Error MessagesIf an error code appears on the display, use the information in Table 9-1 as a troubleshooting guide. If you

Página 109 - Counterpart Limited Warranty

90 Counterpart Counting Scale Indicator9.2 Using the XE and XEH EDP CommandsThe XE and XEH EDP commands can be used to remotely query the Counterpart

Página 110 - For More Information

Appendix 919.3 Continuous Output (Stream) FormatFigure 9-1 shows the default continuous output format sent to a Counterpart port when that port’s STRE

Página 111

92 Counterpart Counting Scale IndicatorB2 Configuration =1 if even parityB3 Dynamic =1 if MODE=NETB4 Dynamic =1 if COZB5 Dynamic =1 if standstillB6 Dy

Página 112

Appendix 93B19 Configuration =000 if secondary DECPNT = 8888800=001 if secondary DECPNT = 8888880=010 if secondary DECPNT = 8888888=011 if secondary D

Comentários a estes Manuais

Sem comentários